本文へスキップ

アナログ MCA, ADC, SDD AMP, 蛍光X線関連の開発、研究を主な業務としています。>Laboratory equipment corporation

見出しHEADLINE

  PRODUCTSschedule


    

ADC2318 is ADC module for MCA.Constituited by Exclusive use of nuclear detective peak detect circuit and ADC. this product have Crucial spec requested to MCA, such as "unmatched accuracy for peak detection" and "fine differencial nonlinerity". those specs are confirmed by pulser at 1 channel.(pulse 1ch.jpg)。

tempreture stability is customed by 100ppm to 10ppm.
requested power has been developed to more ecological as 50mA to 18mA.and half sized.

test report and inspection report is able to order optionaly.

 2318-1.jpg channel test by a pulser.
 2318-2.jpg confirming differencial nonlinerity by sliding pulser.
 2318-3.jpg spectrum of logarismic scale by sliding pulser.
 2318-4.jpg peak detected signal.(1ch blue)
 2318-5.jpg After the peakdetect, start AD convertion immediately. convertion done in
        about 600nSec (2ch light blue) and reset the peakhold. then, allow next input.
 2318-6.jpg AD convertion data processing time it self at MCA is 300nSec.
        so if peaking time is 0.6uSec dead time of MCA can be ignored.




 ○ Temperature stability was 100ppm but 10ppm now.
 ○ Power suply required has minimized by 50mA to 18mA.
 ○ and SIZE!



SPEC ADC resolution:4096ch
AD Converting speed:1.2uSec
input pulse:peaking time of 100ns(Reference value :50ns)
integral nonlinerity :+/−0.01%
differencial nonlinerity :+/−0.63%
tempreture stability:10ppm/℃
input signal:0〜5V、(if add 500Ω at input 0〜10V)
input inpedance:500Ω(0〜5V)、1kΩ(0〜10V)
size:32x20x8
power:+5V 18mA, -5V 9mA, +3.3V 0.13mA (1Kcps at measuring)




which is ideal to test the ADC2318 quicker and understand. board schematic exept ADC module is attached.

test board power:by USB
MCA memory size:4096ch 32bit/ch
through put:about 800Kcps
size(cm):1.5x5.5x9.1 (visitting card size) 

協力企業及び団体
       (敬称略)





過去に参加した展示会等↓